|

MAKING
THE DARKROOM SAFE
A
darkroom should be exactly what the word implies - a room void of any
light. Because panoramic and cephalometric film uses light emitted by
the intensifying screens to expose the image on the film, it is important
that your darkroom not have any lights that will expose your film. The
easiest way to check your darkroom for any light entering from outside
is to turn off all lights in the darkroom, including the safelight.
Allow your eyes to adjust to darkness for a few minutes. Make note of
any areas where you can see light in the darkroom. Some common examples
of light leaks are found are around the door, ceiling tiles or light
fixtures. All areas where you can see visible light entering your darkroom
will impede your film quality. Any other equipment in the darkroom with
operational lights should be relocated to another area if possible or
have the light source covered. Automatic processors with operational
lights are the only exception.
Any light
entering your darkroom can fog the film. Fog can be thought of as background
noise that affects the overall image of the film. The intensity of the
light that hits the film as well as the length of time the film is exposed
to the light can cause varying degrees of fog. A simple test will check
your darkroom for fog conditions. Take a blank sheet of film and place
flat objects, such as coins, on the surface. Allow the film to sit for
two minutes and then develop as you normally would. If you can see the
shape of the object on the film, you have a light source fogging the
film.
SAFELIGHTING
Safelights
provide lighting in the darkroom. But they only provide a short window
of time before they will also fog the film. Safelights allow for approximately
2 1/2 minutes of handling time before fog will develop on the film.
The safelight should have a Kodak GBX-2 filter or equivalent, be placed
a minimum of 4' from the film handling surface and have a safelight
bulb with a maximum of 15 watts. Safelight filters do fade with use
and need to be replaced every three to five years.
MANUAL
PROCESSING/DIP TANKS
Correct
developing of the exposed film can not be overemphasized. Manual processing
requires that you maintain active solutions by replenishing regularly,
using the proper equipment, and following a standard procedure systematically.
A timer
and thermometer are essential to manual processing. Continually monitor
the developer solution temperature to ensure proper chemical activity.
Once the temperature has been determined, an accurate timer is needed
to process the film for the proper length of time. Consult the following
developing time/temperature chart to determine the appropriate chemical
temperature and processing time for the type of film you are using.
MANUAL
PROCESSING RECOMMENDATIONS FOR EXTRAORAL RADIOGRAPHS
|
Film
|
Develop
|
Rinse
|
Fix
|
Wash
|
|
T-Mat
|
4
min @ 80o F (26.5o C)
5 min @ 76o F (24.5o C)
7 min @ 72o F (22.0o C)
8 min @ 68o F (20.0o C)
|
30
sec continuous
agitation @ 60-80o F
(15.5-29.5o C)
|
2-4
min intermittent
agitation @ 60-80o F
(15.5-29.5o C)
|
5
min @ 60-80o F
(15.5-29.5o C)
|
|
X-Omat
RP
|
2.5
min @ 80o F (26.5o C)
4 min @ 76o F (24.5o C)
6 min @ 72o F (22.0o C)
7 min @ 68o F (20.0o C)
|
30
sec continuous
agitation @ 60-80o F
(15.5-29.5o C)
|
2-4
min intermittent
agitation @ 60-80o F
(15.5-29.5o C)
|
5 min @ 60-80o F
(15.5-29.5o C)
|
|
X-Omat
Duplicating
|
2
min @ 80o F (26.5o C)
3 min @ 76o F (24.5o C)
4 min @ 72o F (22.0o C)
5 min @ 68o F (20.0o C)
|
30
sec continuous
agitation @ 60-80o F
(15.5-29.5o C)
|
2-4
min intermittent
agitation @ 60-80o F
(15.5-29.5o C)
|
5
min @ 60-80o F
(15.5-29.5o C)
|
Follow
the instructions on the containers to mix the chemicals properly. Use
the correct amount of replenishing solution to maintain the level as
well as the chemical activity. As a general rule, add eight ounces of
developer daily to the developing solution. You may need to drain some
chemicals out of the tank in order to do this. The solution level must
cover the top of the film hanger. Chemicals should be changed every
3-4 weeks regardless of use since they will break down on their own.
Broken down or old chemicals can not be replenished after
3-4 weeks of use.
Film hangers
should be used to help film avoid making contact with the sides of the
developing tank and other films. Film hangers also aid in film drying.
It is important when loading the film hanger not to scratch the emulsive
surface of the film.
Developing
tanks need one-gallon inserts to process a panoramic size film. Tanks
should have continually running wash water to maintain the developer
temperature and flush away contaminated solutions from the wash tank.
Next
Page
For
technical support, call 800-654-2027 or
e-mail: tech-suppport@pancorp.com.
[PC-1000/Laser
1000] [Literature]
[PC-1000] [Pricing]
[Rental Program]
[Dimensions] [Technical Support] [FAQs]
[Corporate Overview] [Home]
©2001
Panoramic Corporation All Rights Reserved
|